Performance of laser fluorescence at tooth surface and histological section

Performance of laser fluorescence at tooth surface and histological section

Research output: Contribution to journalArticle

View graph of relations

Authors

  • Anahita Jablonski-Momeni
  • David N. J. Ricketts
  • Stefanie Rolfsen
  • Richard Stoll
  • Monika Heinzel-Gutenbrunner
  • Vitus Stachniss
  • Klaus Pieper

Research units

Info

Original languageEnglish
Pages171-178
Number of pages8
JournalLasers in Medical Science
Journal publication dateMar 2011
Journal number2
Volume26
DOIs
StatePublished

Abstract

This study aimed to evaluate a laser fluorescence device (the DIAGNOdent) and a visual classification system (ICDAS-II) for occlusal caries diagnosis. It also aimed to determine whether fluorescence measurements taken at the tooth surface correlate with the fluorescence measurements taken within the body of the lesion. The occlusal surfaces of 100 extracted permanent teeth were examined using ICDAS-II and DIAGNOdent (LF-tooth). Serial sections were made and lesion depth was assessed histologically. DIAGNOdent readings were also taken from the sections (LF-section). There were significant positive strong correlations between ICDAS-II and histology (r(S) = 0.71) and LF-section and histology (r(S) = 0.70), and only moderate correlations between LF-tooth and histology (r(S) = 0.51) and LF-tooth and LF-section (r(S) = 0.60). Diagnostic accuracy for ICDAS-II was generally better than for LF-tooth. While the DIAGNOdent device provides an objective reading for detection and monitoring of carious lesions, using the cut-off ranges previously suggested leads to inferior performance.

Documents

Library & Learning Centre

Contact | Accessibility | Policy