Abstract
This paper describes the initial steps we have taken in establishing CCD based optical-CT as a viable alternative for 3-D radiation dosimetry. First, we compare the optical density (OD) measurements from a high quality test target and variable neutral density filter (VNDF). A modulation transfer function (MTF) of individual projections is derived for three positions of the sinusoidal test target within the scanning tank. Our CCD is then characterized in terms of its signal-to-noise ratio (SNR). Finally, a sample reconstruction of a scan of a PRESAGETM (registered trademark of Heuris Pharma, NJ, Skillman, USA.) dosimeter is given, demonstrating the capabilities of the apparatus.
Original language | English |
---|---|
Article number | 026 |
Pages (from-to) | 195-198 |
Number of pages | 4 |
Journal | Journal of Physics: Conference Series |
Volume | 56 |
DOIs | |
Publication status | Published - 2006 |
ASJC Scopus subject areas
- General Physics and Astronomy