Abstract
Background
Otitis Media with Effusion (OME) is a prevalent childhood condition, affecting up to 90% of children with 10% progressing to a chronic state. It is commonly treated by tympanostomy tube placement. Diagnosis relies on clinical history, otoscopy, audiometry, and tympanometry, but these methods have a high false-positive rate of 15–28%. This highlights the need for an objective, reliable, and non-invasive diagnostic tool to reduce unnecessary surgeries.
Methods
To evaluate the rate of false-positive OME diagnoses in our tertiary hospital, a retrospective database analysis was conducted. To assess the feasibility of advanced imaging, experimental studies were performed ex vivo on human body donors. Sodium chloride solution was intratympanically injected under endoscopic guidance. Hyperspectral imaging (HSI) datasets were acquired using a customized endoscopic system to quantify tissue water content.
Results
An analysis of hospital records identified a false positive rate of 29% (n = 4,564) for OME diagnosis, comparing the number of paracenteses to paracenteses plus tympanostomy tubes. In the ex vivo analysis of HSI, visual evaluation of the Tissue Water Index (TWI) demonstrated sensitivity, specificity, and accuracy of 89%, 78%, and 83%, respectively. The mean TWI of all evaluated tympanic membranes was 50.2 (± 7.1) before injection (control) and increased to 62.8 (± 14.8) after injection.
Conclusion
HSI presents a highly sensitive and user-friendly potential future diagnostic adjunct for diagnosing OME, which has shown its potential in an ex vivo approach. Now it needs in-vivo confirmation of the results as its implementation in clinical routine could contribute to avoiding unnecessary surgical interventions, minimizing patient burden and healthcare costs.
Otitis Media with Effusion (OME) is a prevalent childhood condition, affecting up to 90% of children with 10% progressing to a chronic state. It is commonly treated by tympanostomy tube placement. Diagnosis relies on clinical history, otoscopy, audiometry, and tympanometry, but these methods have a high false-positive rate of 15–28%. This highlights the need for an objective, reliable, and non-invasive diagnostic tool to reduce unnecessary surgeries.
Methods
To evaluate the rate of false-positive OME diagnoses in our tertiary hospital, a retrospective database analysis was conducted. To assess the feasibility of advanced imaging, experimental studies were performed ex vivo on human body donors. Sodium chloride solution was intratympanically injected under endoscopic guidance. Hyperspectral imaging (HSI) datasets were acquired using a customized endoscopic system to quantify tissue water content.
Results
An analysis of hospital records identified a false positive rate of 29% (n = 4,564) for OME diagnosis, comparing the number of paracenteses to paracenteses plus tympanostomy tubes. In the ex vivo analysis of HSI, visual evaluation of the Tissue Water Index (TWI) demonstrated sensitivity, specificity, and accuracy of 89%, 78%, and 83%, respectively. The mean TWI of all evaluated tympanic membranes was 50.2 (± 7.1) before injection (control) and increased to 62.8 (± 14.8) after injection.
Conclusion
HSI presents a highly sensitive and user-friendly potential future diagnostic adjunct for diagnosing OME, which has shown its potential in an ex vivo approach. Now it needs in-vivo confirmation of the results as its implementation in clinical routine could contribute to avoiding unnecessary surgical interventions, minimizing patient burden and healthcare costs.
| Original language | English |
|---|---|
| Article number | 371 |
| Number of pages | 9 |
| Journal | BMC Medical Imaging |
| Volume | 26 |
| Early online date | 29 Jul 2026 |
| DOIs | |
| Publication status | E-pub ahead of print - 29 Jul 2026 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Tympanostomy
- Hyperspectral imaging
- Otoscopy
- Middle ear effusion
- Otitis media with effusion
- Tympanic membrane