T1 - Evaluation of Goldmann applanation tonometry using a nonlinear finite element ocular model
AU - Elsheikh, Ahmed
AU - Wang, Defu
AU - Kotecha, Aachal
AU - Brown, Michael
AU - Garway-Heath, David
N1 - dc.publisher: Springer
PY - 2006
Y1 - 2006
N2 - Goldmann applanation tonometry (GAT) is the internationally accepted standard for intra-ocular pressure (IOP) measurement, which is important for the diagnosis of glaucoma. The technique does not consider the effect of the natural variation in the corneal thickness, curvature and material properties. As these parameters affect the structural resistance of the cornea, their variation is expected to lead to inaccuracies in IOP determination. Numerical Analysis based on the finite element method has been used to simulate the loading conditions experienced in GAT and hence assess the effect of variation in corneal parameters on GAT IOP measurements. The analysis is highly nonlinear and considers the hyper-elastic J-shaped stress–strain properties of corneal tissue observed in laboratory tests. The results reveal a clear association between both the corneal thickness and material properties, and the measured IOP. Corneal curvature has a considerably lower effect. Similar trends have been found from analysis of clinical data involving 532 patients referred to the Glaucoma Unit at Moorfields Hospital, and from earlier mathematical analyses. Nonlinear modelling is shown to trace the behaviour of the cornea under both IOP and tonometric pressure, and to be able to provide additional, and potentially useful, information on the distribution of stress, strain, contact pressure and gap closure.
AB - Goldmann applanation tonometry (GAT) is the internationally accepted standard for intra-ocular pressure (IOP) measurement, which is important for the diagnosis of glaucoma. The technique does not consider the effect of the natural variation in the corneal thickness, curvature and material properties. As these parameters affect the structural resistance of the cornea, their variation is expected to lead to inaccuracies in IOP determination. Numerical Analysis based on the finite element method has been used to simulate the loading conditions experienced in GAT and hence assess the effect of variation in corneal parameters on GAT IOP measurements. The analysis is highly nonlinear and considers the hyper-elastic J-shaped stress–strain properties of corneal tissue observed in laboratory tests. The results reveal a clear association between both the corneal thickness and material properties, and the measured IOP. Corneal curvature has a considerably lower effect. Similar trends have been found from analysis of clinical data involving 532 patients referred to the Glaucoma Unit at Moorfields Hospital, and from earlier mathematical analyses. Nonlinear modelling is shown to trace the behaviour of the cornea under both IOP and tonometric pressure, and to be able to provide additional, and potentially useful, information on the distribution of stress, strain, contact pressure and gap closure.
KW - Tonometry
KW - Intra-ocular pressure
KW - Cornea
KW - Numerical modelling
U2 - 10.1007/s10439-006-9191-8
DO - 10.1007/s10439-006-9191-8
M3 - Article
SN - 0090-6964
VL - 34
SP - 1628
EP - 1640
JO - Annals of Biomedical Engineering
JF - Annals of Biomedical Engineering
IS - 10
ER -