MR-compatible assistance system for biopsy in a high-field-strength system: initial results in patients with suspicious prostate lesions

Stephan Zangos, Andreas Melzer, Katrin Eichler, Cyrus Sadighi, Axel Thalhammer, Boris Bodelle, Renate Wolf, Tatjana Gruber-Rouh, Dirk Proschek, Renate Hammerstingl, Cindy Mueller, Martin G. Mack, Thomas J. Vogl

    Research output: Contribution to journalArticlepeer-review

    74 Citations (Scopus)

    Abstract

    Purpose: To examine the feasibility and safety of magnetic resonance (MR)-guided biopsy by using a transgluteal approach in patients with suspicious prostate lesions by using an MR-compatible robotic system and a 1.5-T MR system.

    Materials and Methods: The study was approved by the institutional review board of University Frankfurt, and informed consent was obtained from each patient. A total of 20 patients (age range, 57.8-71.9 years; mean age, 65.1 years) underwent biopsy in a closed-bore high-field-strength MR system. Biopsy was performed with an MR-compatible pneumatically driven robotic system. T1-weighted gradient-echo fast low-angle shot and T2-weighted true fast imaging with steady-state precession sequences were used to plan and guide the intervention with a transgluteal access on the external planning computer of the assistance system. The system calculated the trajectory and then moved the guiding arm to the insertion point. The cannula was advanced manually, and biopsies were performed with the coaxial technique by using a 15-gauge pencil tip needle. Intervention time, complications, and biopsy findings were documented.

    Results: The MR-compatible robotic system did not interfere with image quality, nor did MR imaging cause dysfunction of the robot. In one patient, the interventionist caused a fail-safe system shutdown. This was due to inadvertent displacement of the guiding arm during cannula insertion. This problem was solved by increasing the displacement threshold. Accurate coaxial cannula biopsy could be performed in all subsequent patients. Sufficient histopathologic assessment was performed in 19 patients. Insufficient material was retrieved in the patient who experienced fail-safe system shutdown. The median intervention time was 39 minutes (23-65 minutes). No procedure-related complications were observed.

    Conclusion: Preliminary results indicate that MR-guided robot-assisted biopsy is feasible and can be performed safely with highly accurate cannula placement. (C) RSNA, 2011

    Original languageEnglish
    Pages (from-to)903-910
    Number of pages8
    JournalRadiology
    Volume259
    Issue number3
    Early online date1 Mar 2011
    DOIs
    Publication statusPublished - Jun 2011

    Keywords

    • Transgluteal biopsies
    • Antigen levels
    • Cancer
    • Feasability
    • Device
    • Interventions
    • Validation
    • Navigation
    • Placement

    Fingerprint

    Dive into the research topics of 'MR-compatible assistance system for biopsy in a high-field-strength system: initial results in patients with suspicious prostate lesions'. Together they form a unique fingerprint.