Abstract
Purpose To explain the observed considerable loss of signal and contrast when Thiel-embalmed human cadavers are imaged using clinical magnetic resonance imaging (MRI) sequences, especially those based on spin-echo MRI. Materials and Methods All cadavers were imaged with a medical 1.5T scanner using standard MRI sequences. Dual angle B1+ magnitude mapping and electromagnetic (EM) simulations that characterize the radiofrequency (RF) penetration in a male human body model (HBM) were carried out for a range of tissue conductivities. Results The EM simulations show that RF penetration issues begin to affect the image quality for values of electrical conductivity as low as 2.6 S/m. The electrical conductivity values of the embalming fluids were found to be within the range of 5-10.6 S/m, thus strongly suggesting that the observed loss in signal and contrast is due to diminished RF penetration inside the cadavers. Furthermore, it was demonstrated that gradient-echo (GRE)-based MRI sequences perform better than spin-echo (SE)-based sequences, as they are less susceptible to imperfections in the flip angle that are inevitably present when imaging Thiel cadavers. Conclusion: The diminished signal and contrast observed when imaging Thiel-embalmed human cadavers may be attributed to the high conductivity of the embalming liquids. J. Magn. Reson. Imaging 2014;39:576-583. © 2013 Wiley Periodicals, Inc. Copyright © 2013 Wiley Periodicals, Inc.
Original language | English |
---|---|
Pages (from-to) | 576-583 |
Number of pages | 8 |
Journal | Journal of Magnetic Resonance Imaging |
Volume | 39 |
Issue number | 3 |
Early online date | 1 Jan 1970 |
DOIs | |
Publication status | Published - Mar 2014 |
Fingerprint
Dive into the research topics of 'MRI of thiel-embalmed human cadavers'. Together they form a unique fingerprint.Profiles
-
Melzer, Andreas
- Neuroscience - Professor and Director of IMSAT (Teaching and Research)
Person: Academic