Optimized deployment of heat activated surgical staples using thermography

Y. Ng, C. Song, D. McLean, S. M. Shimi, T. G. Frank, A. Cuschieri, P. A. Campbell

    Research output: Contribution to journalArticlepeer-review

    8 Citations (Scopus)

    Abstract

    We have developed a suture analogue, suitable for use in laparoscopic surgery, in the form of a staple constructed from NiTi shape memory alloy (SMA). Closure of the staple is effected by resistive heating via a (50–100 ms) pulse of electrical current (5 A). In order to optimize deployment protocols and minimize thermal collateral damage to tissue, the heat sink effect of the electrical contact rails, as well as the presence of contact hotspots must be considered. Here, we have employed high-resolution thermal imaging to observe the dynamic temperature distributions in SMA staples as a function of the pulse parameters. This has facilitated process optimization and also provided data from which to validate computational finite-element models of the heat transport phenomena. © 2003 American Institute of Physics.
    Original languageEnglish
    Pages (from-to)1884-1886
    Number of pages3
    JournalApplied Physics Letters
    Volume83
    Issue number9
    DOIs
    Publication statusPublished - Sept 2003

    Keywords

    • Infrared imaging
    • Surgery
    • Biothermics
    • Temperature distribution
    • Shape memory effects
    • Finite element analysis
    • Laparoscopic surgery