Performance of a 50 MHz annular array based imaging system

J.A. Brown, C Morton-Démoré, G.R. Lockwood, F.S. Foster

    Research output: Chapter in Book/Report/Conference proceedingOther chapter contribution

    4 Citations (Scopus)

    Abstract

    The performance of a high frequency annular array based imaging system is described. Radiation patterns generated by the array + beamformer were evaluated by imaging a point target immersed in a water bath. The lateral image resolution at a depth of 5.5 mm was 96 microns and the secondary lobes were suppressed 60 dB below the main lobe. Preliminary images of CD-1 mice were also generated using the array system. The images were displayed with a dynamic range of up to 60 dB.
    Original languageEnglish
    Title of host publicationProceedings - IEEE Ultrasonics Symposium
    Pages760-763
    Number of pages4
    Volume1
    DOIs
    Publication statusPublished - 1 Jan 2004

    Fingerprint

    Dive into the research topics of 'Performance of a 50 MHz annular array based imaging system'. Together they form a unique fingerprint.

    Cite this