Performance of laser fluorescence at tooth surface and histological section

Anahita Jablonski-Momeni, David N. J. Ricketts, Stefanie Rolfsen, Richard Stoll, Monika Heinzel-Gutenbrunner, Vitus Stachniss, Klaus Pieper

    Research output: Contribution to journalArticlepeer-review

    36 Citations (Scopus)

    Abstract

    This study aimed to evaluate a laser fluorescence device (the DIAGNOdent) and a visual classification system (ICDAS-II) for occlusal caries diagnosis. It also aimed to determine whether fluorescence measurements taken at the tooth surface correlate with the fluorescence measurements taken within the body of the lesion. The occlusal surfaces of 100 extracted permanent teeth were examined using ICDAS-II and DIAGNOdent (LF-tooth). Serial sections were made and lesion depth was assessed histologically. DIAGNOdent readings were also taken from the sections (LF-section). There were significant positive strong correlations between ICDAS-II and histology (r(S) = 0.71) and LF-section and histology (r(S) = 0.70), and only moderate correlations between LF-tooth and histology (r(S) = 0.51) and LF-tooth and LF-section (r(S) = 0.60). Diagnostic accuracy for ICDAS-II was generally better than for LF-tooth. While the DIAGNOdent device provides an objective reading for detection and monitoring of carious lesions, using the cut-off ranges previously suggested leads to inferior performance.

    Original languageEnglish
    Pages (from-to)171-178
    Number of pages8
    JournalLasers in Medical Science
    Volume26
    Issue number2
    DOIs
    Publication statusPublished - Mar 2011

    Keywords

    • ICDAS II
    • Laser fluorescence
    • Histological sections
    • Occlusal caries detection
    • in vitro
    • ICDAS-II
    • Reproducibility
    • Validity
    • Systems
    • Device

    Fingerprint

    Dive into the research topics of 'Performance of laser fluorescence at tooth surface and histological section'. Together they form a unique fingerprint.

    Cite this