Skip to main navigation Skip to search Skip to main content

Too Many Shades of Grey: Photometrically and Spectrally Mismatched Targets and Backgrounds in Printed Acuity Tests for Infants and Young Children

  • Guillermo Vivas-Mateos
  • , Iain A. T. Livingstone
  • , Ruth Hamilton
  • , Arsalan Cheema
  • , Mario Ettore Giardini (Lead / Corresponding author)

    Research output: Contribution to journal › Article › peer-review

    35 Downloads (Pure)

    Abstract

    Purpose: Acuity tests for infants and young children use preferential looking methods that require a perceptual match of brightness and color between grey background and target spatial average. As a first step in exploring this matching, this article measures photometric and colorimetric matches in these acuity tests.

    Methods: The luminance, uniformity, contrast, and color spectra of Teller Acuity Cards, Keeler Acuity Cards for Infants, and Lea Paddles under ambient, warm, and cold lighting, and of grey-emulating patterns on four digital displays, were measured. Five normal adults’ acuities were tested at 10 m observationally.

    Results: Luminance and spectral mismatches between target and background were found for the printed tests (Weber contrasts of 0.3% [Teller Acuity Cards], −1.7% [Keeler Acuity Cards for Infants], and −26% [Lea Paddles]). Lighting condition had little effect on contrast, and all printed tests and digital displays met established adult test luminance and uniformity standards. Digital display grey backgrounds had very similar luminance and color whether generated by a checkerboard, vertical grating, or horizontal grating. Improbably good psychophysical acuities (better than −0.300 logMAR: (logarithm of the minimum angle of resolution)) were recorded from adults using the printed tests at 10 m, but not using the digital test Peekaboo Vision.

    Conclusions: Perceptible contrast between target and background could lead to an incorrectly measured, excessively good acuity. It is not clear whether the luminance and spectral contrasts described here have clinically meaningful consequences for the target patient group, but they may be avoidable using digital tests.

    Translational Relevance: Current clinical infant acuity tests present photometric mismatches that may return inaccurate testing results.
    Original languageEnglish
    Article number12
    Number of pages14
    JournalTranslational Vision Science and Technology
    Volume9
    Issue number12
    DOIs
    Publication statusPublished - 6 Nov 2020

    UN SDGs

    This output contributes to the following UN Sustainable Development Goals (SDGs)

    1. SDG 3 - Good Health and Well-being
      SDG 3 Good Health and Well-being

    Keywords

    • visual acuity
    • Teller Cards
    • Keeler Acuity Cards
    • Lea Paddles
    • Cardiff Cards
    • Peekaboo Vision

    ASJC Scopus subject areas

    • Ophthalmology

    Fingerprint

    Dive into the research topics of 'Too Many Shades of Grey: Photometrically and Spectrally Mismatched Targets and Backgrounds in Printed Acuity Tests for Infants and Young Children'. Together they form a unique fingerprint.

    Cite this