Skip to main navigation Skip to search Skip to main content

An Application of Optical Coherence Tomography in Intraocular Pressure Measurement and Effect Evaluation

  • Daiyuan Xie

    Student thesis: Master's Thesis › Master of Philosophy

    Abstract

    To provide an auxiliary tool for the screening and treatment of IOP-related diseases like glaucoma, a novel non-contact intraocular pressure (IOP) estimation technique based on the surface area of the corneal depression region in optical coherence tomography (OCT) images was proposed in this study. The study first reviewed the clinical significance of glaucoma and the importance of IOP measurement, analysed the limitations of traditional measurement methods, and introduced the development background and advantages of OCT technology. Experiments were conducted using a porcine eye model, and corneal image data at different pressure levels were acquired using a commercial OCT device. Two surface area estimation procedures were developed to quantify the surface area by fitting a depression profile curve, and a univariate IOP regression model was constructed based on this.

    The experimental results showed a significant correlation between the estimated IOP and the measured values ( 𝑅² = 0.595, 𝑝 = 0.005 ), which verified the feasibility and potential of the method. Further analyses combined with biomechanical principles explored the relationship between corneal deformation and IOP and the factors affecting estimation accuracy. The study also pointed out the limitations of the current method in terms of consistency between different samples and adaptability to asymmetric deformations. For this reason, it is proposed to introduce multifactor modelling, artificial intelligence techniques to optimize the estimation accuracy, improve the image processing algorithm, and expand the sample size to enhance the model generalisation ability in the future.

    In summary, this paper constructs an IOP estimation process based on the structural parameters of OCT images, which provides a preliminary practice for non-invasive, image-driven IOP assessment assistance. The method has a broad clinical application prospect, especially suitable for glaucoma remote screening and personalised medicine, which has important research value and promotion significance.
    Date of Award2025
    Original languageEnglish
    Awarding Institution
    • University of Dundee
    SupervisorChunhui Li (Supervisor) & Ale Aranceta-Garza (Supervisor)

    Cite this

    '